doi: 10.7763/IJAPM.2013.V3.192
Construction of the Optical Delay Line for the Optical Coherence Tomography
- 1 the Physics Department, Faculty of Science, King Mongkut’s Institute of Technology Ladkrabang, Bangkok 10520, Thailand
- 2 the Department of Physics, Chung Yuan Christian University, Chungli 32023, Taiwan ROC
Abstract
We have constructed the compact optical delay line for the axial scanning of time-domain optical coherence tomography. The delay line contains the retro-reflector, and the inclined reflection mirror, as well as the scanning mirror. This delay line is performed by the low-coherence reflectometer with a scanning speed of 400 Hz. The dimension of the delay line is 2 cm × 2 cm. We have achieved the scanning range of about 3 mm within the scanning mirror’s vibration angle of approximately 10°.
Keywords
- Imaging technique
- interferometer
- optical delay line
- reflectometer
How to Cite
Kitsakorn Locharoenrat and I-Jen Hsu, "Construction of the Optical Delay Line for the Optical Coherence Tomography," International Journal of Applied Physics and Mathematics, vol. 3, no. 2, pp. 137-139, 2013. https://doi.org/ 10.7763/IJAPM.2013.V3.192
Copyright & License
Copyright © 2013 by the authors. This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited (CC BY 4.0).